Arthroplasty
Metasul metal-on-metal THRs have high rate of revision and incidences of radiolucent lines
This report has been verified
by one or more authors of the
original publication.
Bone Joint J. 2013 Jul;95-B(7):881-6. doi: 10.1302/0301-620X.95B7.31067
399 hips undergoing total hip replacement were randomized to receive either Metasul metal-on-metal (MoM), metal-on-conventional polyethylene (MoP), or ceramic-on-polyethylene (CoP) bearings using a cemented triple-tapered polished femoral component (MS-30). The study aimed to compare the three types of bearings in clinical and radiological outcomes. The results taken at 2,5,and 7 years revealed that MoM group had the lowest Harris Hips scores, higher rates of revision, and higher incidence of radiolucent lines in unrevised hips.